The Reliability of a Posterior-to-Anterior Spinal Stiffness Measuring System in a Population of Patients With Low Back Pain
-
A schematic diagram of the PA stiffness system. The Polhemus Liberty electromagnetic tracking system monitors displacement, whereas force is measured with an inline force transducer.
A schematic diagram of the PA stiffness system. The Polhemus Liberty electromagnetic tracking system monitors displacement, whereas force is measured with an inline force transducer.
-
Force vs displacement plots of 5 loading and unloading cycles on the third lumbar SP of a patient in the LBP trial. The nonlinear range is less than 30 N. The more linear 55- to 75-N force range was u
Force vs displacement plots of 5 loading and unloading cycles on the third lumbar SP of a patient in the LBP trial. The nonlinear range is less than 30 N. The more linear 55- to 75-N force range was used to calculate stiffness.
-
A scatterplot of stiffnesses measured for the 2 successive tests of PA stiffness taken 5 minutes apart. Slp_A is the calculated slope for the first test, averaged from the last 4 cycles of indentationA scatterplot of stiffnesses measured for the 2 successive tests of PA stiffness taken 5 minutes apart. Slp_A is the calculated slope for the first test, averaged from the last 4 cycles of indentation. Slp_B is the average slope found for the second test [R2 = 0.628; ICC(3,1), 0.790].